The EBC is designed for use together with 3
rd
party castors and the LINAK OpenBus™
control system, including Jumbo Care. The EBC - Electronic brake control - can replace
traditional (mechanical) central locking systems for castors on hospital and nursing home
beds or medical applications like e.g. shower equipment. This new system offers an
increased freedom of application design as there is no need for a mechanical connection
between the castors.
4. EBC - Electronic Brake Control System (MEDLINE
®
CARELINE
®
)
Usage:
Operating temperature: 5°C to 40°C
Storage temperature: - 10°C to 50°C
Approved according to IEC60601-1:2005 3
rd
edition approved, ANSI / AAMI ES60601-1:2005, 3
rd
edition, and CAN / CSA-22.2 No 60601-1:2008
Recommendations:
Always use locking mechanism and O-ring on cables.
If any open sockets they must be fitted with blind plugs to ensure IP Degree.
Removing or adding any OpenBus™ cables is not allowed when the control box is powered (hot plugging).
Before installation/service
Stop the application.
Remove battery power cable and OpenBus™ connection then possible castor connection.
Service system.
After installation/service
Reconnect castor connection, then OpenBus™ connection, battery power cable, then control box mains.
LINAK A/S only delivers the OpenBus™ system and is not responsible for any products other than LINAK products (i.e. products from 3
rd
party suppliers or the compliance of such products with the LINAK OpenBus™ system).
...to be continued
The massage motor can be added to all kinds of couches and tables, chairs or
beds for treatment and examination. It enables comfort, relaxation and tension
release for patients and clients. The massage motors are directly connected to the
actuator port at the control box – no extra wiring at the application, simple and
easy mounting.
5. Massage Motor Medical (MEDLINE
®
CARELINE
®
)
Usage:
Compatibility: CB6S OBMe, CB16 OBF, (CB20 pending) MJB006-0x to be used for OpenBus™ impulse drive
Duty cycle: 10 %, 30 min. max.
Operating temperature: +5°C to +40°C
Storage temperature: -10°C to +50°C
Relative humidity: 20 % to 80 % at +30°C
Atmospheric pressure: 700 to 1060 hPa
Approvals: Medical approvals to be determined