CARELINE
®
)
The HB80 hand control has an optimised ergonomic design shaped for the hand. The handset is suitable for all
kinds of MEDLINE and CARELINE applications such as hospital beds, patient lifts, treatment and examination
couches etc.
The HB80 handset is available in versions with up to 10 or 12 activation buttons.
Usage:
• Usage temperature: 5º C to 40º C
• Storage temperature: -10º C to +50º C
• Compatibility: CB6, CB8, CB9, CB12, CB14, CBJ 1/2, CBJ-H, CBJ-C, CB6S, CB16 and CB20.
• Approvals: IEC60601-1:2005 3
rd
edition, ANSI/AAMI ES60601-1:2005 3
rd
edition and CAN/CSA-22.2 No 60601-1:2008.
The HB86 version has a shorter distance between the buttons and cannot be approved according to EN IEC60601-2-52
Application Environment 4 for care beds used in Domestic area (or EN1970). HB80 is designed and tested according to
IPX6 washable in conformity with EN60601-2-52 cl. 201.11.6.6.101 (Machine washable medical beds).
The HB80 must hang vertically from it’s hook during the washing process.
In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way that the
cable’s own weight does not strain the coil during the washing process.
In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way
that the cable’s own weight does not strain the coil during the washing process.
Recommendation:
• Unless otherwise specified or agreed by LINAK - the handset is only intended to be used on LINAK systems.
• When changing handsets for OpenBus™ systems the power must be switched off.
• It is recommended to check the handset and cable for damage and holes made by violent handling before washing the application or at least once
a year.
• It is recommended to have a parking place for the handset on the application, where the customer ensures that the handset does not fall off.
• Clean the handset regularly to ensure good hygiene standards.
• When a defective HB80 is replaced, check that the new HB80 has exactly the same specification and functionality.
• Do not submerge the handset under water.
For handsets with magnet:
• If handsets with magnet are attached to a smooth surface, a movement or twisting of the cable, for example during transport, can cause the
handset to move and result in damage if the cable is squeezed.
• The force of the magnet depends on the thickness of the lacquering, the lacquering type, stickers, steel thickness etc.
It is the responsibility of the customer to verify that the holding force on the application is acceptable.
• It is the responsibility of the user/operator to evaluate any possible risk caused by use of permanent magnets.
Warnings!
• Do not sit or lie on the handset. It can cause unintended movement of the application.