Page 74 of 108
The HB30 hand control is designed for better user experience and ergonomic fit for the hands
of caregivers. The compact size ensures one hand operation. The HB30 is especially suitable for
patient lifts and other MEDLINE
®
and CARELINE
®
applications like couches, tables and chairs
for treatment and examination.
13. HB30 (MEDLINE
®
CARELINE
®
)
Usage:
• Usage temperature: 5° C to 40° C
• Storage temperature -10° C to +50° C
• Compatibility: Analogue JUMBO Systems
Analogue JUMBO systems with diode and OpenBus JUMBO versions
All OpenBus control boxes
• Approvals: IEC60601-1:2005 3
rd
edition approved, ANSI / AAMI
ES60601-1:2005 3
rd
edition and CAN/CSA-22.2
No 60601-1:2008 pending
The HB30 is biocompatibility tested and approved according to DS/EN ISO 10993-5:2009, biological evaluation of
medical devices - part 5: Tests for in vitro cytotoxicity. It is a demand for hand-held devices for patient lifts.
The HB30 has a compact design and therefore it cannot be approved according to EN IEC60601-2-52
(Application Environment 4 for care beds used in Domestic areas (or EN1970)).
How to identify the cables:
Each cable has a label for easy identification of item number and for which control box it is intended.
...to be continued
How to mount a cable:
Step 1:
Mount the cable lock and fix it to the slot
marked in the picture.
Step 2:
Fix the cable tab on the hand control’s front
side first. Push in and twist a bit to fix the tab
(see picture fit A into B).
Step 3:
Fix the tab on the back as well by pushing.